49 / 72 page ![]() Data Sheet ADE7758 Rev. E | Page 49 of 72 Gain Calibration Using Line Accumulation Step 2: Select Phase A, Phase B, or Phase C for a line period measurement with the FREQSEL[1:0] bits in the MMODE register (0x14). For example, clearing Bit 1 and Bit 0 selects Phase A for line period measurement. Gain calibration is used for meter-to-meter gain adjustment, APCF or VARCF output rate calibration, and determining the Wh/LSB, VARh/LSB, and VAh/LSB constant. Step 3: Set up ADE7758 for line accumulation by writing 0xBF to LCYCMODE. This enables the line accumulation mode on the xWATTHR, xVARHR, and xVAHR (0x01 to 0x09) registers by setting the LWATT, LVAR, and LVA bits, LCYCMODE[0:2] (0x17), to Logic 1. It also sets the ZXSEL bits, LCYCMODE[3:5], to Logic 1 to enable the zero-crossing detection on all phases for line accumulation. Additionally, the FREQSEL bit, LCYCMODE[7], is set so that FREQ (0x10) stores the line period. When using the line accumulation mode, the RSTREAD bit of LCYCMODE should be set to 0 to disable the read with reset mode. Select the phase for line period measurement in MMODE[1:0]. Step 0: Before performing the gain calibration, the APCFNUM/ APCFDEN (0x45/0x46) and VARCFNUM/ VARCFDEN (0x47/0x48) values can be set to achieve the correct impulses/kWh, impulses/kVAh, or impulses/kVARh using the same method outlined in Step 4 in the Gain Calibration Using Pulse Output section. The calibration of xWG/xVARG/xVAG (0x2A through 0x32) is done with the line accumulation mode. Figure 82 shows the steps involved in calibrating the gain registers using the line accumulation mode. Step 1: Clear xWG, xVARG, and xVAG. Step 4: Set the number of half-line cycles for line accumulation by writing to LINECYC (0x1C). FREQUENCY KNOWN? NO YES STEP 0 SET APCFNUM/APCFDEN AND VARCFNUM/VARCFDEN STEP 1 STEP 2 CLEAR xWG/xVAR/xVAG STEP 3 SET LYCMODE REGISTER STEP 4 SET ACCUMULATION TIME (LINECYC) STEP 5 SET MASK FOR LENERGY INTERRUPT STEP 6 SET UP SYSTEM FOR ITEST,VNOM,PF = 1 STEP 7 READ FREQ[11:0] REGISTER STEP 8 RESET STATUS REGISTER STEP 9 READ ALL xWATTHR AND xVAHR AFTER LENERGY INTERRUPT STEP 9A CALCULATE xWG STEP 9B CALCULATE xVAG STEP 10 WRITE TO xWG AND xVAG CALIBRATE WATT AND VA @ PF = 1 STEP 11 SET UP TEST SYSTEM FOR ITEST,VNOM, PF = 0, INDUCTIVE STEP 12 RESET STATUS REGISTER STEP 13 READ ALL xVARHR AFTER LENERGY INTERRUPT STEP 14 CALCULATE xVARG STEP 15 WRITE TO xVARG STEP 16 CALCULATE Wh/LSB, VAh/LSB, VARh/LSB END SELECT PHASE FOR LINE PERIOD MEASUREMENT CONFIGURE FREQ[11:0] FOR A LINE PERIOD MEASUREMENT Figure 82. Gain Calibration Using Line Accumulation |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |