14 / 24 page ![]() 14 ENABLE THE QCL INSTRUMENT Press the Enable button on the front panel. The button will glow blue only when current is delivered to the load. There is a standard 2-second delay before the button illuminates and current flows. VI SCAN 1. On the Control screen, press to access the VI scan screen. 2. Press the Current Start value to select the current at which to start the scan. Turn the Adjustment knob to change it. Figure 19. Scan Screen 3. Press the Current Stop value to select the current at which to stop the scan. 4. Press the Step Size value to select the current interval at which to take a measurement. The maximum interval is 10% of full scale. Minimum step is 1% of full scale. 5. Press the Time Interval value to select the amount of time to allow the current to settle. For the best data accuracy, use a Time Interval of 2 seconds or more. The minimum time interval is 0.1 seconds. 6. Insert the USB flash drive in the USB slot on the front panel. 7. Press the green Run button to start the scan. When finished, data is saved to the USB flash drive with the filename viscanxx.txt, where xx starts at 00 and increments. 8. To stop mid-scan, press the red Stop button to abort the scan. Any data collected during the aborted scan is saved to the USB flash drive. NOTES: • The step interval time is approximate. In the scan files, time intervals are slightly more than the step interval indicated on the instrument. Each interval builds on the previous one. • Should the Step Size interval not finish at the Stop Current, the reading is still taken and recorded. For example, if the Start Current is 250 mA and the Stop Current is 1.04 A with a Current Step of 250 mA, a reading is taken at 1000 mA (1 A) and 1040 mA (1. 04 A) but not at 1250 mA (1.25 A). • After a VI scan is completed, the output is disabled and the setpoint is zeroed. SAVE OR RECALL SETTINGS Press to access the Save/Recall Screen. Under Save Settings to, press the blue text to toggle through the Profile options, select one, and then press the Save icon. The following variables are saved: OPERATION SETTINGS VI SCAN SETTINGS Setpoint VI Start Current Current Limit VI Stop Current Max Supply Voltage VI Step Size Cable Resistance VI Time Interval Display Brightness Beeper Status Figure 20. Save / Recall Screen Under Recall Settings from, press the blue text to toggle through the saved Profile options, and then press the Recall icon to recall the saved settings. Select Factory to restore the settings to the factory defaults. |
|
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |