69 / 199 page ![]() AS7058 Functional description Datasheet • PUBLIC • Document Feedback DS001077 • v2-00 • 2025-Sep-24 69 / 199 / Standby control signal Function Timing control registers Functionality Standby Enable 13 Controls and enables common mode voltage of ECG amplifier time1_en13 time2_en13 Required for ECG operation only Standby Enable 14 Controls and enables ECG amplifiers time1_en14 time2_en14 Required for ECG operation only The timings which are defined in the timing registers for each control signal define the start of a block prior to the actual measurement start. The timing registers allow for exact control of the startup timing behavior of all relevant blocks and it makes sure that all system blocks are active and properly settled when a measurement is started. Right after the measurement is finished and the measurement data is written to the FIFO memory of AS7058 all Standby Enable signals are disabled again and the connected blocks are powered down. Important to mention is that Standby Enable 1 – Standby Enable 5 signals do have only one timing control register (time_enX). Each register is linked to the 32 kHz clock domain of the device which results into a 31.25 µs multiplication factor for these registers like it is also indicated in Figure 62. The Standby Enable 6 – Standby Enable 14 control signals have two timing control registers. The time1_enX timing registers are also linked to the 32 kHz clock domain of the device which results into a 31.25 µs multiplication factor for these registers. The second timing control register time2_enX which is also linked to each of the Standby Enable 6 – Standby Enable 14 registers is connected to the 2 MHz clock domain of AS7058 which results in a 1 µs multiplication factor for these registers. Only one of the two-timing registers for a Standby Enable signal can be active at a time and the priority is controlled with the register content. In case a value greater than zero is written to time1_enX register the register content of time2_enX register is ignored and the timing is calculated and executed based on the register content of time1_enX register. In case the time1_enX register is set to zero, the content of register time2_enX is used for the Standby Enable signal timing based on the 1 µs clock domain. |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |